Troubleshooting Checks Action
9 Was the test strip exposed to open air for a long
period of time?
Repeat the control test with properly stored strips.
9 Was the test strip bottle capped tightly?
If the bottle was not capped tightly, the strips could
be humidified. Replace the test strips.
9 Was the meter functioning well?
You can use control solution to verify the meter’s
functions. (Chapter 2)
9 Is the control solution expired or contaminated?
Replace with new control solution to check the
performance of SMBG system.
9 Were test strips and control solutions stored in
cool, dry places?
Repeat the control test with properly stored strips
or control solutions.
9 Did you follow the testing steps properly?
Read Chapter 2 "Control Solution Testing" and test
again. Stop using the meter if you continue to
obtain the inaccurate results. Call Consumer
Services Toll-Free: 866-994-3345 (Eastern Time,
Mon-Fri 9:00AM~5:00PM). During non-service
hours, please contact your healthcare provider.