Message What it means? What to do?
The test result is lower than 20
mg/dL.
This may require immediate treatment
according to your healthcare
professional’s recommendations.
Although this message could be due to
a test error, it is safer to treat first and
then do another test.
The “Ht” and thermometer icon
appears. Temperature is too high,
outside the required range of 10°C -
40°C (50°F - 104°F). This alerts users
that an incorrect result may occur if
the test continues.
Relocate the meter to a location with
temperature between 10°C - 40°C
(50°F - 104°F).
The “Lt” and thermometer icon
appears. Temperature is too low,
outside the required range of 10°C -
40°C (50°F - 104°F). This alerts users
that an incorrect result may occur if
the test continues.
Relocate the meter to a location with
temperature between 10°C - 40°C
(50°F - 104°F).