The "Ht" and thermometer icon appears.
Temperature is too high, outside the required
range of 50°F ~104°F (10°C ~40°C). This alerts
users that an incorrect result may occur if the
test continues.
A
ction: Relocate the meter to a location with
temperature between 50°F ~104°F (10°C
~40°C).
The "Lt" and thermometer icon appears.
Temperature is too low, outside the required
range of 50°F ~104°F (10°C ~40°C). This
alerts users that an incorrect result may
occur if the test continues.
A
ction: Relocate the meter to a location with
temperature between 50°F ~104°F (10°C
~40°C).